Air Science Purair RX Ductless Fume Hood
This enclosure is a Class 1 product that meets USP 800 requirements for non-sterile compounding procedures. USP 800 guidelines specify that hazardous materials must be compounded within the proper work environment associated with both sterile and non-sterile products and preparations. This enclosure is designed to protect the user and the environment from hazardous powders and particulates generated on the work surface. Option to purchase a second HEPA filter (7045) to convert this to a double HEPA hood.
*Base stand not included*
With Canopy and Phenolic Resin Work Surface
2 ft Product Dimensions: W 24" × D 27" × H 35"
3 ft Product Dimensions: W 36" × D 27" × H 35"
4 ft Product Dimensions: W 48" × D 27" × H 35"
- A strategically placed front cover window shows the installed filter part number and installation date to encourage timely filter replacement
- Rear baffle provide smooth horizontal airflow pattern. Removable for easy cleaning
- Standard 6-in diameter exhaust connection port to allow for outside ducting. USP-800 compliant
- Hoods include standard 60 W lights to fully illuminate the work surface
- When closed, the cabinet sash protects the contents from inadvertent external contact, and better isolates the air within. The sash is easy to open and close and features a self-locking feature
- Electronic controls and displays include switched for the blower and low AirFlow alarm. Ergonomics, safety and aesthetics all come together with the 10° pitch of the face
- The chemical resistant epoxy coated steel frame adds mechanical strength
- The 99.5 % effective electrostatic pre-filter is accessible from inside the chamber to contain the release of any particulates that it traps. The pre-filter can be changed while the unit is operating to prevent operator exposure to chemical vapors
- Electrical cords and cables are safely routed into the cabinet through 1 5/8 diameter port on the back and side walls
- A continuous air velocity monitoring system alerts the operator upon unacceptable values
- Built-in blower
- Authorized personnel may set the centrifugal fan motor speed as desired
- The internal work surface is fitted with a standard epoxy resin surface
- Filter access keys prevent unauthorized removal or accidental exposure to dirty filters