On December 24, the office will have reduced hours and be closed on December 25.

Air Science Purair RX Ductless Fume Hood

Air Science Purair RX Ductless Fume Hood

This enclosure is a Class 1 product that meets USP 800 requirements for non-sterile compounding procedures. USP 800 guidelines specify that hazardous materials must be compounded within the proper work environment associated with both sterile and non-sterile products and preparations. This enclosure is designed to protect the user and the environment from hazardous powders and particulates generated on the work surface. Option to purchase a second HEPA filter (7045) to convert this to a double HEPA hood.

 

*Base stand not included*

 

With Canopy and Phenolic Resin Work Surface

 

2 ft Product Dimensions: W 24" × D 27" × H 35"

3 ft Product Dimensions: W 36" × D 27" × H 35"

4 ft Product Dimensions: W 48" × D 27" × H 35"

  • A strategically placed front cover window shows the installed filter part number and installation date to encourage timely filter replacement
  • Rear baffle provide smooth horizontal airflow pattern. Removable for easy cleaning
  • Standard 6-in diameter exhaust connection port to allow for outside ducting. USP-800 compliant
  • Hoods include standard 60 W lights to fully illuminate the work surface
  • When closed, the cabinet sash protects the contents from inadvertent external contact, and better isolates the air within. The sash is easy to open and close and features a self-locking feature
  • Electronic controls and displays include switched for the blower and low AirFlow alarm. Ergonomics, safety and aesthetics all come together with the 10° pitch of the face
  • The chemical resistant epoxy coated steel frame adds mechanical strength
  • The 99.5 % effective electrostatic pre-filter is accessible from inside the chamber to contain the release of any particulates that it traps. The pre-filter can be changed while the unit is operating to prevent operator exposure to chemical vapors
  • Electrical cords and cables are safely routed into the cabinet through 1 5/8 diameter port on the back and side walls
  • A continuous air velocity monitoring system alerts the operator upon unacceptable values
  • Built-in blower
  • Authorized personnel may set the centrifugal fan motor speed as desired
  • The internal work surface is fitted with a standard epoxy resin surface
  • Filter access keys prevent unauthorized removal or accidental exposure to dirty filters